The standard EN ISO 16409:2006 + A1:2010 contains an testof the Filament Retention.
The brush is moved to a hole 80 times. The brush must lose no filament.
The device includes a linear motion mechanism with motor. The start and end position can be teached. The number of cycles is freely programmable.
The drive allows a constant speed for a defined test path.
2 Limit switches in the device allow checking the speed.
The jigs allow the rapid fixation of different interdental brushes. Various brackets are attached.
Spezification
Dimension (H,W,D) ca. | 320, 260, 250 mm | |
Weight | 12 kg | |
Cycle Speed | 0.25 - 1 Hz | |
Distance | 0 - 25 mm | |
Power Supply | 115/230 (50/60 Hz) | |
EN ISO 13849-1 Kategorie 3 und PL d. |
Ordering
Device Durability Test 308-030 9.058 €
Measure Plate 308-020 1.247 €
Scope of Delivery
- 1 Device + Plate
- 1 Power Cord CEE 16A (5 m)
- 1 Operating Intruction
Conditions of Delivery
Delivery: EXW
Shipping time: by arrangement
Technical changes reserved.
The illustrations may differ from the actual equipment.
All data without obligation and not-binding.