- Simple, intuitive operation through procedure templates, minimum
Operator training
- Dust and splash-proof touchscreen
- Modern Design & Technology: linear guidance technology with stepper motor
ensure high precision in distance and speed
- Independent, fully operational system,
requires no computer and no software installation
-Verification Function
The standard EN ISO 16409:2006 + A1:2010 contains a test of the Stem Retention.
Insert the manual interdental brush head through the selected hole. Place the clamp on the stem. The stem
should be secured at approximately the midpoint of the interdental brush head. Pull the clamp at a pulling
speed of (20 ± 2) mm/min to remove the stem from the interdental brush handle.
The device includes a linear motion mechanism. The start and end position can be teached. The speed is freely programmable.
Spezification
Force Sensor | 50 N 0,2%, 2 KHz | |
Dimension (H,W,D) ca. | 420, 420, 330 mm | |
Weight | 25 kg | |
Speed | 20 - 60 mm/min | |
Distance | 0 - 50 mm | |
Power Supply | 115/230 (50/60 Hz) | |
Safety | EN ISO 13849-1 PL 4 / e |
Ordering
Interdental brush Stem Retention Test 308-040 12.820 €
Scope of Delivery
- 1 Device + Plate
- 1 Force Sensor 50 N / 2000 Hz
- 1 Power Supply 1.625A
- 1 Operating Intruction
Conditions of Delivery
Delivery: EXW
Shipping time: by arrangement
Technical changes reserved.
The illustrations may differ from the actual equipment.
All data without obligation and not-binding.